S0002001

FACSPrestoCD4abs%counter-start kit&acces


FACSPrestoCD4abs%counter-start kit&acces
Indicative Price : 9,610.00 USD

General Description:
BD FACSPresto Point-of-Care (PoC), CD4 [absolute, % and Hb] counter with starter kit and accessories.

Technical Specifications:
Functionally complete table top counter, portable for CD4 absolute count with start kit. Mainly used for ART initiation and monitoring of adults, adolescents and children living with HIV/AIDS.

·Single-test processing automatic CD4 enumerator, portable (6kg).

·Technology: 3-color Imaging cytometry with fluorescent labelled antibodies to count CD4 and % in whole blood. Imaging spectrometer for absorbance measurement of total Hemoglobin.

·Provides CD4 count (Cells/ul), CD4 percentage (%) and Hemoglobin (g/dL) quantitative results.

·Accommodates dedicated prefilled single-use testing cartridges.

·Sample volume for analysis: less than 20ul of capillary or venous whole blood.

·Range, approximately
CD4 absolute: 50-4000 cells/ul
CD4 %: 5%-60%
Hemoglobin concentration: 2 g/dL-20g/dL

·Test reading time: 22 minutes for first sample. After that 4 minutes per sample for batched samples.

·Throughput of samples per operator per day (8 hours working day). 80 samples.

·Large color touchscreen display informs:
Device operational status and on-going analysis
Measured results per patient ID, with date and time
System errors or malfunctions, incl. insufficient sample volume
Status built-in battery

·Results menu will allow data filtration for printing and export via USB port.

·Factory calibrated and built-in auto-self-test, each time the device is switched-on.

·Built-in printer able to print test results, patient ID, Instrument QC, Reagent QC, expiration date, operator name, lab location, date, time, software version.

·The system has capacity to store 12000 test results with all the information associated with the runs.

·Built-in on board Instrument QC in the instrument and reagent QC feature in the cartridge.

·Design of prefilled cartridge insertion port, eliminates sample cross-contamination.

·Cartridges shelf-Life: 12 months.

·Ambient operating conditions ranges:
Operational temperature: 10 to 40°C
Relative humidity: 10 to 95%

·Built-in rechargeable battery, 6 hours when fully charged, approximately. 40 test.

·Power requirements: Built-in battery and 100-240 (A/C) at 50-60 Hz.

·System has built-in voltage surge protection. No need for UPS.

·Robust, easy to clean and maintain.

·Equipment with 3 year warranty.

Manufacture product reference: 651000, 658212 and 658860 (BD International)

Items supplied with:
1x 100 Cartridges
1x 100 Finger Stick Sample collection kit:
1x 100 Lancets
1x 100 alcohol swabs
1x 100 cotton gauze
1x 100 band aids
1x 100 transfer pipets
1x Power supply and 4 adaptor cords for different regions
1x Workstation
1x Equipment cover
5x Rolls of thermal printer paper
1x USB drive
1x Solar charger kit
1x Car charger adaptor
1x Power generator
1 x Instructions for use in English, French, Spanish
1x Quick Reference Guide

Items required, but not supplied:
·Disposable gloves, anti-septic swabs, absorbent paper, protective glasses, timer, container for biohazard disposables and disinfectant (equipment and surfaces).

·Consumables for venous blood collection: EDTA blood collection tubes, vacuum tube needle, needle holder, micropipettes and centrifuge.

Packaging:
First box: BD FACSPresto PoC, CD4 [absolute, % and Hb] counter with starter kit, accessories and instructions for use.
Second box: BD FACSPresto CD4 PoC, Solar charger kit and charger adaptor
Third box: BD FACSPresto CD4 PoC, Car battery

Shelf life:
12 months from the date of manufacturing; do not use beyond its expiry date.

Storage and transportation:
Operating temperature between 10-40ºC, protected against humidity and direct sunlight.

General recommendations:
·Contains hazardous materials in limited quantities. For more information refer to manufacturers Material Safety Data Sheet (MSDS).

·Follow strictly manufacturer instructions of the test procedure.

·Do not substitute reagents from other kits/batches.

·Do not freeze reagents.

Weight and volume:
First box
Estimated weight: 10.50kg
Estimated volume: 68.45cdm
Second box:
Estimated weight: 15kg
Estimated volume: 102.31cdm
Third box:
Estimated weight 1kg
Estimated volume: 2,655cdm