S0001518
Refrig/freezer, lab,2-8C/-20C, 180L/40L
Refrig/freezer, lab, 2-8C/-20C, 180L/40L
Combined Refrigerator and Freezer, lab,
Operating temperature (refrigerator): 2-8°C
Operating temperature (freezer): -20°C
Refrigerator gross volume: 185, 188 or 225 L
Freezer gross volume: 97 or 100 L
Quantity: 1 Unit
Indicative Price 1,500.00 USD
General description:
Refrig/freezer, lab, 2-8C/-20C, 180L/40L
Combined Refrigerator and Freezer, lab,
Operating temperature (refrigerator): 2-8°C
Operating temperature (freezer): -20°C
Refrigerator gross volume: 185, 188 or 225 L
Freezer gross volume: 97 or 100 L
Quantity: 1 Unit
Technical specifications:
Product Description: Upright refrigerator for storage of chemicals and reagents in clinical laboratory with following minimum specifications:
1.Compression type, CFC-free refrigerant.
2.Insulation material: polyurethane, CFC-free.
3.Fan-cooled for even distribution of air in the cabinet.
4.Stainless steel structure or its equivalent or inner chamber PS plate depending on the refrigerator model.
5.Refrigerator gross volume: 185, 188 or 225 L
6.Freezer gross volume: 97 or 100 L
7.Easily adjustable shelves.
8.Lockable door, solid.
9.Electronic temperature control (refrigerator): 2°C to 8°C.
10.Electronic temperature control (freezer): up to -20°C.
11.Accuracy, regardless of the load: +/- 1°C.
12.Ambient operating temperature acceptable range: Lower 10°C, upper 30 °C.
Temperature monitoring:
1.External digital display with actual interior temperature, minimal graduation 0.1°C.
2.Electronic temperature recording device: includes data logger (optional or standard, depending on the refrigerator model).
3.Audio and visual alarm system indicates unsafe temperatures (optional or standard, depending on the refrigerator model).
4.Battery back-up for audio and visual alarm system, and temperature recording device optional or standard, depending on the refrigerator model).
Other requirements:
1.Fitted with integrated castors.
2.Minimum compressor starting voltage compressor starting voltage: +-10%.
3.Meeting quality standard ISO 8187 / EN 28187 or similar.
4.Meeting safety standards: EMI 89/336EEC, 73/23/EEC and 93/68/EEC code AB1 or Similar.
5.Power requirements: 220 V / 50 Hz or 110/60 Hz (if requested).
6.Power consumption: approx. 125 to 430 W or more energy efficient.
Supplied with Automatic voltage regulator:
A microprocessor-controlled spike and surge protection and protection against disturbances.
1-Nominal output voltage: 220 V / 50 Hz, Single phase, 110/60 Hz (If requested), Single Phase.
2-Accepted input range: 160-280 VAC for 220 V/50 Hz system and 82-159 volt for 110/60 Hz (If requested), Single Phase.
3-Output accuracy: ≤ +/- 10 %.
4-Response time: <15 ms or 50 s (depending on the model)
5-LED display showing connected/disconnected status, voltage fluctuation and load as % of nominal current or a digital display showing temperature, voltage, electricity (depending on the refrigerator model).
6-Permissible overload: Permissible overload: 1000 % within 60 ms protection.
7-Electronic fuse disconnects and reconnects automatically (optional depending on the model).
8-kVA rating matches power consumption of the refrigerator.
Warranty:
The Refrigerator is to be covered by a two-year replacement warranty in the event of any component failure arising from defective design, materials or workmanship.
Spare parts:
Manufacturer will publish a list of spare parts recommended for 5 years of operation (in operation manual, website or as a list to be sent with refrigerator). Manufacturer will ensure supply of spare parts for a minimum of 5 years from the time of cessation of the last production of equipment.
Note: Spare parts supply is not mandatory during the warranty period.
Materials:
Restricted materials: The product and its constituent components, does not contain lead (except in batteries), mercury, cadmium, hexavalent chromium, polybrominated biphenyls (PBB) or polybrominated biphenyl ethers (PBDE).
Other restricted materials: The appliance and its constituent components, does not contain lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls (PBB) or polybrominated biphenyl ethers (PBDE).
Instructions for use:
The refrigerator can be used in wide range of laboratory services from basic laboratory to reference centre.
Instructions for use, preventive maintenance and troubleshooting are supplied in English as a standard with the refrigerator (French and Spanish instructions available depending on the refrigerator model).
Packaging and Labelling:
Refrigerator will be packed with suitable protective packaging for the international shipment such as; packing foam, corrugated carton, non-fumigation pallets, pallet straps. The packaging will be compliant to restricted material requirements mentioned above.
Labelling:
Compliance with EAN 128 bar code requirements.
Weight - Volume:
Estimated Unit Weight in Kg (including its packaging): 139.33 kg
Estimated Unit Volume in M3 (including its packaging): 1.08 m3