S0002479

Mains Ref & Frz Haier HBCD-90 E003/097


Mains Powered, Ice-lined Combined Refrigerator & Waterpack Freezer for Storage of Vaccines
Haier HBCD-90 E003/097
Ambient operating temperature: Min. +5°C, Max. +43°C
Net vaccine storage volume: 30 L
Supplied with voltage regulator and temperature monitoring device.
Indicative Price 1,733.00 USD
Disclaimer: The images shown are for illustration purposes only and may not be an exact representation of the product. “No Generic Photo Available" means that there is not an image available for this product.

Mains Powered Combined Vaccine Refrigerator and Water pack Freezer
Equipment model: HBCD-90
PQS code: E003/097
Manufactured by: Qingdao Haier Biomedical Co.,Ltd in China
Volume category: Vaccine storage capacity: 30L

Technical specifications
PQS Prequalified Intermittent Mains Powered Combined Refrigerator and Water-Pack Freezer for storage of vaccines at +2 ˚C to+8˚C and freezing of water packs.
220-240 Volt or 50Hz / 60Hz, 1ph
Operating temperature: Min and Max rated ambient temperature: +5 °C and +43°C
UIFPF- Grade A
Refrigerant: R600a, (45 +40) grams

To be supplied with:
Standalone Voltage stabilizer for 220-240V 50/60 Hz nominal input/output:
Option 1: Extended type Haier HVS-1000U, E007/009, with input voltage regulation range of 110V-278V/50/60Hz in compliance with PQS_E007_VS01.6
Option 2: Extended type Sollatek SVS04-22E, E007/004, with input voltage regulation range of 110V-278V/50/60Hz
Power lead and plug.
With PQS prequalified 30-day temperature logger.
RTMS – Optional/upon request
The door / lid fitted with a lock and two keys and 4 denser moulded PP engineering plastics baskets (2 in refrigerator and 2 in freezer for ice pack).
Warranty: 3- year replacement warranty as specified in the LTA.

NOTES:
1.UNICEF CO or partners should notify the voltage & frequency to ensure correct choice of freezer and voltage regulator.
2.Provide the preferred power plug type compatible with the electricity socket standard in the country
3.One set of spare parts is recommended but not compulsory for every 10 units of mains powered appliances.
4.For Technical Guidance: Contact with UNICEF Supply Division, Immunization Technology Center, Cold Chain Unit for technical questions. CPHHQ-SD-Cold Chain Unit < coldchainunit@unicef.org >

Useful links for reference:
1.UNICEF Cold Chain Technical Support:
2.UNICEF Supply Catalogue:
3.WHO-PQS Website:
4.GAVI CCEOP Technology Guide:
5.WHO Vaccine Management Handbook:
Related Products