S0002499
SDD Ref. Haier HTC-120 E003/116
Mains Powered, Ice-lined Refrigerator for Storage of Vaccines
Haier HBC-240 E003/115
Ambient operating temperature: Min. +5°C, Max. +43°C
Net vaccine storage volume: 200 L
Supplied with voltage regulator and temperature monitoring device.
Indicative Price 3,450.00 USD
Solar Direct Drive Vaccine Refrigerator with no battery
Equipment model: HTC-120
PQS code: E003/116
Manufactured by: Qingdao Haier Biomedical Co.,Ltd in China
Volume category: >/= 60, <120L
Vaccine storage capacity:100L
Technical specifications
PQS prequalified Solar direct drive refrigerator with no battery for storage of vaccines at +2 ˚C to+8˚C in areas without any electricity, or where there is less than eight hours of reliable electricity over a typical day.
Solar radiation reference period of 3.5 kWh/m²/day and hot zone (+43°C) test temperatures.
Operating temperature: Min and Max rated ambient temperature: +5 °C and +43°C
UIFPF: Grade A
Refrigerant: R600a, 65 grams
To be supplied with:
PQS prequalified 30-day temperature logger.
RTMS – Optional/upon request
The door / lid fitted with a lock and two keys, 3 denser wired shelves
With photovoltaic system (solar panels) and all necessary support structure, fasteners, cables, connectors, etc. for roof mounting as a standard.
Warranty for equipment: 3- year replacement warranty.
NOTES:
1.One set of spare parts is recommended but not compulsory for every 10 units of SDDs.
2.For Technical Guidance: Contact with UNICEF Supply Division, Immunization Technology Center, Cold Chain Unit for technical questions. CPHHQ-SD-Cold Chain Unit < coldchainunit@unicef.org >