S0372020

Tube,asp/feed,CH12,L125cm,ster,disp


Tube, aspirating / feeding, CH12, length 125 cm, sterile, disposable.
Indicative Price 0.16 USD
Disclaimer: The images shown are for illustration purposes only and may not be an exact representation of the product. “No Generic Photo Available" means that there is not an image available for this product.

General Description:
Tube, aspirating/feeding, CH12, length 125cm, sterile, disposable

Technical Specifications:
Sterile enteral feeding/aspirating catheter for single use.
Single channel tube.
Proximal end with cup connector, conical tip, allowing the tube to be connected to devices such as feeding syringes (catheter tip) or suction pump system.
Open distal end with four side windows.
Tube with markings at 40, 50, 60 and 70cm from the distal end.
Diameter expressed in Charriere, French gauge.
Length expressed in cm.
Colour code/ external diameter: Visible on cup connector.
Material: Polyvinyl chloride (PVC).
Size selected: Diameter CH12, length approximately 125cm
Disposable
Sterile
Initial sterilisation method: Ethylene oxide gas

Packaging and labelling:
Primary packaging: Unit of use.
One (1) aspirating/feeding tube in an individual sterilised
peel pack.

Secondary packaging: Protected unit.
One (1) box of 50 aspirating/feeding tubes.

Over packaging: Packaging unit.

Weight/Volume/Dimensions:
Estimated weight: 0.022kg
Estimated volume: 0.232cdm

Instructions for use:
Sterile aspirating/feeding tube:
Device used for aspirating liquids and / or gases from stomach, duodenum and small intestine.
Device used for feeding child with feeding syringe conical tip
Marking on the tube: 40, 50, 60, 70cm from the distal end. Correspond to the following regions:
cardia, stomach, duodenum, jejunum.
The size has been chosen as being the most commonly used.

Remember to order when needed:
S0366010 - Syringe, feeding, 50ml, conical tip, sterile

Conditions for stock:
Avoid storage at extreme temperatures and humidity levels.
Check the integrity of each unit before use.
Single use material, supplied in sterile packaging.
Do not use if the packaging is damaged.

Safety process:
The aspirating / feeding tube is for single use only.
When inserting a gastric tube, use clean techniques.
Collect and destroy through incineration in a controlled environment.